Пресс-релиз популярных книг
.
Авторы: 111 А Б В Г Д Е Ж З И Й К Л М Н О П Р С Т У Ф Х Ц Ч Ш Щ Э Ю Я
Книги: 164 А Б В Г Д Е Ж З И Й К Л М Н О П Р С Т У Ф Х Ц Ч Ш Щ Э Ю Я
На сайте 111 авторов, 92 книг, 72 статей, 5913 глав.
3.6 MEASUREMENT OF SOUND POWER
The sound pressure in the vicinity of a noise source is generally dependent
on the surroundings. The sound pressure level will be different for the same
noise source, for example, if the source is located indoors or if it is located
outdoors. The sound pressure will be different if the source is placed in a
room with acoustically reflective surfaces or if the room surfaces are highly
absorbent for sound. In contrast, the sound power is generally independent
of the surroundings. For this reason, information about the sound power
spectrum for a noise source is important to the designer interested in noise
control.
There is no ‘‘acoustic wattmeter’’ available for direct measurement of
sound power, however. The sound power must be inferred (or calculated)
from measurements of sound pressure or sound intensity and other appropriate
quantities, such as surface area, reverberation time, etc.
Acoustic Measurements 51
FIGURE 3-8 Personal noise dosimeter. Data can be stored in several locations to
allow the monitoring of multiple inputs. (By permission of Casella CEL Instruments
Ltd.)
Copyright © 2003 Marcel Dekker, Inc.
There are three broad classes of environment used in connection with
sound power determination: (a) reverberant field, (b) direct or anechoic
field, or (c) the actual environment to which the noise source is exposed
(in-situ survey). The national and international standards for sound power
measurement in a reverberant room include ANSI S1.31, ANSI S1.32, and
ANSI S1.33 (Acoustical Society of America, 1986a,b,c) and ISO 3741, ISO
3742, ISO 3743 (International Organization for Standardization, 1986a,b,c).
The corresponding standards for an anechoic room include ANSI S1.35
(Acoustical Society of America, 1979a) and ISO 3745 (International
Organization for Standardization, 1986d). The survey method is covered
by ANSI S1.36 (Acoustical Society of America, 1976b) and ISO 3746
(International Organization for Standardization, 1986e).
A reverberant room is a room in which the acoustic energy from sound
reflected from the room surfaces (reverberant field) is much larger than the
energy transmitted directly from the noise source to a receiver (direct field),
as discussed in Sec. 7.2. All surfaces in a reverberant room are highly reflective
or have a very low surface absorption coefficient. A reverberant room
may be used to determine sound power by either comparison with a calibrated
noise source or by direct measurement of sound pressure.
An anechoic room is a room in which practically all of the acoustic
energy striking the surfaces of the room is absorbed. Because the energy
reflected from the room surfaces is negligible in an anechoic room, the
energy transmitted directly from the source to the receiver is predominant.
Measurements in an anechoic room may be used to determine the directional
characteristics (directivity factor or directivity index) of the noise
source. One modification of the anechoic room is the semi-anechoic room,
in which the floor surface is highly reflective, but the other surfaces in the
room are highly absorptive. An alternative arrangement is to make measurements
outdoors on a reflective surface, such as a parking lot.
In situations where the noise source cannot be moved into a reverberant
or anechoic room, the sound power may be determined from measurements
taken in situ, with appropriate adjustments made for surrounding
surfaces and environment background noise.
3.6.1 Sound Power Measurement in a
Reverberant Room
The determination of the sound power in a reverberant room requires that
the diffuse or reverberant sound field in the room is much larger than the
direct sound field. This requirement results in a practically uniform value of
the acoustic energy density and the acoustic pressure in the room.
52 Chapter 3
Copyright © 2003 Marcel Dekker, Inc.
The volume of the reverberant room should be such that the wavelength
of the sound waves is much smaller than the dimensions of the room.
The minimum volume for the room should meet the following condition:
V р3_Ю3 ј 9рc=f Ю3 (3-6)
The quantity c is the speed of sound in the air in the room, and f is the
octave band (or 1/3-octave band) center frequency for the lowest-frequency
band considered in the measurements. For air at 300K
(c ј 347:2m=s ј 1139 fps) and a frequency of 125 Hz, the corresponding
wavelength is as follows:
_ ј р347:2=125Ю ј 2:78m р9:11 ftЮ
The minimum reverberant room volume for this condition may be found
from Eq. (3-6).
V р9Юр2:78Ю3 ј 193m3 р6810 ft3Ю
If the room dimensions (height, width, and length) are in the commonly
used ratio 1 : 1:5 : 2:5, the room dimensions must be at least 3.72m (12.2 ft)
high, 5.58m (18.3 ft) wide, and 9.30m (30.5 ft) long. Room dimension ratios
of 1 : 21=3 : 41=3 or 1 : 1:260 : 1:587 have also been used (Broch, 1971). The
volume of the equipment being tested should not exceed 0.01V, where V is
the volume of the room.
The room surfaces in a reverberant room should have surface absorption
coefficients (Sabine absorption coefficient) that are less than about 0.06.
The surface absorption coefficient is discussed in Sec. 7.1.
The sound field in the region near the walls of the room will not be
quite uniform or diffuse, so it is good practice to locate the microphones
such that none are nearer than the smaller value of 1
2 _ or 1m (39 in) from the
walls. For case of a wavelength of 2.78m given in the previous example
(1
2 _ ј 1:36 m), the microphone should be located at least 1m from the walls.
If an array of microphones is used, at least three microphones should be
included in the array. If a single microphone is used, measurements should
be taken at three or more positions around the noise source. The positions
should be spaced at a distance that is the larger of 3р1
2 _Ю or 3 m, where _ is
the wavelength of the lowest-frequency sound to be measured.
The noise source should not be placed at the center of the room,
because many of the resonant modes of the room would not be excited by
the noise source in this position. The noise source is usually placed near the
room wall at a distance not less than the major dimension of the source.
Acoustic Measurements 53
Copyright © 2003 Marcel Dekker, Inc.
3.6.1.1 Comparison Method
The sound power may be measured by comparison of the measured sound
pressure level in a reverberant room with the sound pressure level of a
reference (calibrated) sound source at the same location. A reference
sound power source was originally designed by a committee of the
American Society of Heating, Refrigeration, and Air Conditioning
Engineers (ASHRAE) in the 1960s (Baade, 1969). Reference sound power
sources are commercially available with calibration accuracies of 0:5dB
for frequencies between 200 and 4000 Hz and 1:0 dB between 100 and
160 Hz and between 5 and 10 kHz.
Several microphones arranged in an array in the room or a single
movable microphone may be used to measure the sound pressure level in
the reverberant field. To ensure that the microphones are in the diffuse field,
the distance between the microphones and the surface of the noise source dm
should meet the following condition (Beranek and Veґ r, 1992, p. 92):
20 log10р1:25dmЮ LW;cal _ Lp;cal (3-7)
The quantity LW;cal is the sound power level of the reference source, and
Lp;cal is the measured sound pressure level produced by the calibrated sound
power source.
The experimental procedure for determining the sound power level for
a noise source, using the comparison method in a reverberant room, is as
follows. First, the energy-averaged sound pressure level in each frequency
band Lp is measured with the test source in operation. Secondly, the test
source is removed and the reference sound source is placed in the same
location, and energy-averaged sound pressure level in each frequency
band Lp;cal is measured with the calibrated reference source in operation.
Thirdly and finally, the sound power level of the test source LW is calculated
from the measured data:
LW ј Lp ю рLW;cal _ Lp;calЮ (3-8)
The values of the sound power level for the calibrated reference source
LW;cal are supplied by the manufacturer of the calibrated source.
If the reverberant field is much larger than the direct sould field, Eq.
(7-17), which relates the sound power level and sound pressure level, may be
written in the following form:
Lp ј LW ю 10 log10р4=RЮ ю 0:1 (3-9)
The quantity R is the room constant, defined by Eq. (7-13). The value of the
room constant remains constant when the test source is replaced by the
calibrated source:
54 Chapter 3
Copyright © 2003 Marcel Dekker, Inc.
Lp;cal ј LW;cal ю 10 log10р4=RЮ ю 0:1 (3-10)
If we subtract Eq. (3-10) from Eq. (3-9), we obtain the expression given by
Eq. (3-8).
3.6.1.2 Direct Method
If the room constant were known, Eq. (3-9) could be used to determine the
sound power level directly from sound pressure level measurements. If the
acoustic energy density associated with the reverberant sound field is much
larger than that associated with the direct sound field, Eq. (7-14) may be
written as follows:
4W
4 ј
p2
_oc
(3-11)
If the room is highly reverberant, or if the average surface absorption coefficient
__ is small, the room constant from Eq. (7-13) may be written as
follows:
R ј
__So
1 _ __ _ __So (3-12)
The quantity So is the total surface area of the room.
The reverberation time Tr may be used to determine the average surface
absorption coefficient for the room surfaces. The reverberation time,
adjusted for standing wave effects, is given by Eq. (7-34):
Tr ј
55:26V
ca
1 ю
_
2d
_ _
(3-13)
The quantity V is the volume of the room, c is the speed of sound in the air
in the room, and a is the number of absorption units, given by Eq. (7-30).
The quantity _ ј c=f is the wavelength of the sound at the band center
frequency, and d ј 4V=So is the mean free path for the sound waves in
the room. For small values of the surface absorption coefficient, the number
of absorption units may be approximated by the following, according to Eq.
(7-30),
a ј So ln
1
1 _ __
_ _
_ __So (3-14)
By comparing Eqs (3-12) and (3-14), we observe that the room constant and
the number of absorption units are approximately equal.
R _ a _ __So (3-15)
Acoustic Measurements 55
Copyright © 2003 Marcel Dekker, Inc.
If we make the substitutions from Eq. (3-15) into Eq. (3-13) for the
reverberation time, the following result is obtained:
R ј
55:26V
cTr
1 ю
Soc
8Vf
_ _
(3-16)
If we make the substitution for the room constant from Eq. (3-16) into Eq.
(3-11), the following result is obtained for the acoustic power:
W ј
55:26V
4Tr
1 ю
Soc
8Vf
_ _
p2
_oc2 (3-17)
We may convert Eq. (3-17) to ‘‘level’’ form as follows. First, introducing
the reference quantities, we have the following:
W
Wref ј р p=pref Ю2рV=Vref Ю
рTr=Tref Ю
1 ю
Soc
8Vf
_ _
13:816р pref Ю2Vref
TrefWref_oc2 (3-17a)
where Vref ј 1m3 and Tref ј 1 sec. If we take log base 10 of both sides of
Eq. (3-17a) and multiply both sides by 10, we obtain the final result needed
to determine the sound power level of a noise source from sound pressure
level measurements in a reverberant room:
LW ј Lp ю 10 log10рV=VrefЮ _ 10 log10рTr=TrefЮ ю 10 log10 1 ю
Soc
8Vf
_ _
ю 10 log10
13:816р pref Ю2Vref
TrefWref_cc2
" #
(3-18)
For an ideal gas, the last term in Eq. (3-18) may be written in the
following form:
_oc2 ј
Po
RT р_RTЮ ј _Po ј _Po;ref рPo=Po;ref Ю (3-19)
The quantity Po is atmospheric pressure, _ is the specific heat ratio
р_ ј 1:40 for air), and Po;ref ј 101:325 kPa (14.696 psia). If we substitute
the numerical values for the reference quantities, we obtain:
13:816р pref Ю2Vref
TrefWref_oc2 ј р13:816Юр20 _ 10_6Ю2р1Ю
р1Юр10_12Юр1:40Юр101:325 _ 103ЮрPo=Po;ref Ю ј
0:03896
рPo=Po;ref Ю
10 log10
13:816р pref Ю2Vref
TrefWref_oc2
" #
ј 10 log10
0:03896
рPo=Po;ref Ю
_ _
ј _10 log10рPo=Po;refЮ _ 14:1
56 Chapter 3
Copyright © 2003 Marcel Dekker, Inc.
With these substitutions, Eq. (3-18) may be written in the following form:
LW ј Lp ю 10 log10рV=VrefЮ _ 10 log10рTr=TrefЮ ю 10 log10 1 ю
Soc
8Vf
_ _
_ 10 log10рPo=Po;refЮ _ 14:1
(3-20)
To ensure that the diffuse or reverberant sound field predominates at
the microphone location, the distance between the microphones and the
surface of the noise source dm should meet the following condition
(Beranek and Veґ r, 1992, p. 93):
dm 3
V
cTr
_ _1=2
р3-21)
Example 3-2. A reverberant room has dimensions of 6m (19.69 ft) by 10m
(32.81 ft) by 4m (13.12 ft) high. The measured reverberation time for the
room is 3.50 seconds. The air in the room is at 300K (278C or 808F) and
101.3 kPa (14.7 psia), at which condition the speed of sound is 347.2 m/s
(1139 fps). The measured sound pressure level in the 500 Hz octave band due
to the noise from pump in the room is 65 dB. Determine the sound power
level for the pump in the 500 Hz octave band.
The surface area of the room is as follows:
So ј р2Юр10 ю 6Юр4Ю ю р2Юр10Юр6Ю ј 128 ю 120 ј 248m2 р2669 ft2Ю
The volume of the room is:
V ј р10Юр6Юр4Ю ј 240m3 р8476 ft3Ю
The value of the following quantity for a frequency of 500 Hz may be
calculated:
Soc
8Vf ј р248Юр347:2Ю
р8Юр240Юр500Ю ј 0:0897
The sound power level may now be found from Eq. (3-20):
LW ј 65 ю 10 log10р240Ю _ 10 log10р3:50Ю ю 10 log10р1 ю 0:0897Ю
_ 0 _ 14:1
LW ј 65 ю 23:80 _ 5:44 ю 0:37 _ 14:1 ј 69:6dB
The minimum distance of the microphone from the surface of the
pump is given by Eq. (3-21):
Acoustic Measurements 57
Copyright © 2003 Marcel Dekker, Inc.
dm р3Ю р240Ю
р347:2Юр3:50Ю
_ _1=2
ј 1:333m р4:37 ftЮ
3.6.2 Sound Power Measurement in an Anechoic
or Semi-Anechoic Room
The acoustic power generated by a noise source may also be measured in an
environment such that the direct acoustic field is much larger than the
reverberant field. This situation may be achieved in an anechoic or semianechoic
room or outdoors away from any reflecting surfaces, such as buildings,
walls, etc. The directivity characteristics of the noise source may be
measured in an anechoic room, as discussed in Sec. 3.6.4.
In an anechoic chamber, the room surfaces are treated with acoustic
material such that the surface absorption is practically 100%. The floor of a
semi-anechoic room is highly reflective, but the walls are highly absorptive,
as in the case of an anechoic chamber.
An array of microphones or a single microphone moved to various
positions may be used for making the measurements. The microphone measurement
positions may be located at the same distance a from the center of
the noise source on a spherical or hemispherical surface. The radius a of the
sphere or hemisphere should be at least twice the largest dimension of the
source or four times the height of the source, whichever is larger. The
microphone measurement position should be farther than a distance dm ј 1
4 _ from any room surfaces, where _ is the wavelength corresponding to the
lowest octave band center frequency considered in the measurements.
If the direct acoustic field is the only contributor to the acoustic pressure,
the sound power Wj radiated from the noise source through an area Sj
is related to the acoustic intensity Ij and measured acoustic pressure pj as
follows:
Wj ј IjSj ј
Sjp2j
_oc
(3-22)
The total sound power radiated from the noise source is the sum of the
power radiated through the entire surface around the source:
W ј
X
j
Wj (3-23)
The determination of the sound power may be made somewhat more
convenient by dividing the total surface area of the sphere or hemisphere
into Ns equal surface areas.
58 Chapter 3
Copyright © 2003 Marcel Dekker, Inc.
Sj ј 2_a2=Ns (for a hemisphere) (3-24)
Sj ј 4_a2=Ns (for a sphere) (3-25)
The surface area of the surface bounded by angles _1 and _2, for
example, is illustrated in Fig. 3-9. The area is given by:
S2 ј
р_2
_1 р2_a sin _Юрa d_Ю ј 2_a2рcos _1 _ cos _2Ю (3-26)
Let us consider the case for a hemispherical area or radius a, and use Ns ј 10 equal areas. The top ‘‘cap’’ will be one area рS1Ю, and each of the next
‘‘rings’’ will each be divided into three areas. The surface area of the ‘‘cap’’,
with _o ј 08, is as follows:
S1 ј 2_a2р1 _ cos _1Ю ј р1=10Юр2_a2Ю (3-27)
If we solve for the angle _1, we obtain the following:
cos _1 ј 1 _ 0:100 ј 0:900 and _1 ј 25:848 ј 0:4510 rad
Using Eq. (3-26), we may obtain the following values for the other angles:
cos _2 ј cos _1 _ 0:300 ј 0:600 and _2 ј 53:138 ј 0:9273 rad
cos _3 ј cos _2 _ 0:300 ј 0:300 and _3 ј 72:548 ј 1:2661 rad
cos _4 ј cos _3 _ 0:300 ј 0:000 and _4 ј 90:008 ј 1
2 _ rad
The microphone should be placed at the geometrical centroid of the
surface area segment, as shown in Fig. 3-9. The centroid for the ‘‘cap’’ is
directly at the top of the hemisphere (__1 ј 08Ю. The angle locating the centroid
of the ‘‘band’’ areas may be found from the following expression:
Acoustic Measurements 59
FIGURE 3-9 Coordinates for determining the microphone locations on a measuring
surface of radius a. The small circles denote the points at the centroid of the area
segments.
Copyright © 2003 Marcel Dekker, Inc.
рcos _1 _cos _2Ю__2 ј
р_2
_1
_ sin_ d_ р3-28)
__2 ј рsin_2 _sin_1Ю _ р_2 cos _2 __1 cos _1Ю
cos _1 _cos _2
(3-29)
If we substitute the numerical values for a 10-microphone system, the following
value is obtained for the centroid of the first ‘‘band’’ area:
__2 ј
sinр53:138Ю_sinр25:848Ю_Ѕр0:9273Ю cosр53:138Ю
_р0:4510Ю cosр25:848Ю_
р0:900_0:600Ю
__2 ј 0:7121 rad ј 40:808
We may repeat the calculations for the other ‘‘band’’ areas to obtain
the following result:
__3 ј 1:1016 rad ј 63:128
__4 ј 1:4196 rad ј 81:348
The vertical distance from the equator (floor for a hemisphere) yj and
the horizontal distance from the vertical axis xj for the microphone locations
may be found, as follows:
y2=a ј cos__ ј 0:757 and x2=a ј sin__2 ј 0:653
y3=a ј cos__3 ј 0:452 and x3=a ј sin__3 ј 0:892
y4=a ј cos__4 ј 0:151 and x4=a ј sin__4 ј 0:989
The specific locations for 10 microphones, according to ISO3744, are shown
in Fig. 3-10.
For the case in which the measuring surface subdivisions are equal in
area, the total sound power may be found from Eqs (3-22) and (3-23), using
the sound pressure measurements:
W ј _IjSj ј
So_p2
j
_ocNs
(3-30)
The quantity So is the total surface area (So ј 2_a2 for a hemisphere; So ј 4_a2 for a sphere). For a spherical surface, 20 measurements could be used,
for example. The 10 microphone locations below the equator would be at
the same distance below the equator as those above the equator for the
hemispherical surface with 10 microphone locations.
Example 3-3. The sound pressure level measurements given in Table 3-2
were obtained in a semi-anechoic room around a motor. The overall dimen-
60 Chapter 3
Copyright © 2003 Marcel Dekker, Inc.
Acoustic Measurements 61
FIGURE 3-10 Microphone locations for a hemispherical measurement surface,
using 10 measurement points (open circles). The closed circles denote locations for
an additional 10 microphones for improved accuracy, particularly for sources having
nonsymmetrical sound radiation characteristics.
TABLE 3-2 Data for Example 3-3
Point Elevation, y, m Angle, _ Lp, dB
1 1.250 0.08 86.0
2 0.9375 41.48 81.5
3 0.9375 41.48 82.4
4 0.9375 41.48 81.3
5 0.5625 63.38 70.9
6 0.5625 63.38 72.9
7 0.5625 63.38 68.0
8 0.1875 81.48 79.3
9 0.1875 81.48 78.5
10 0.1875 81.48 80.1
Copyright © 2003 Marcel Dekker, Inc.
sions of the motor are 500mm(19.69in) long by 300mm(11.81 in) wide and
300mm (11.81in) high. The microphones were located at a distance of
1.250m (49.21 in) from the center of the motor. The air in the room was
at 248C (297.2Kor 75.28F) and 101.6 kPa (14.74 psia), for which the properties
are sonic velocity c ј 345:6m/s (1134 fps); density _o ј 1:191 kg/m3
(0.0744 lbm=ft3), and _oc ј 411:6rayl. Determine the overall sound power
level for the motor.
The surface area for the measurement hemisphere may be calculated:
So ј р2_Юр1:250Ю2 ј 9:817m2 р105:7ft2Ю
The sumof the squares of the acoustic pressure may be calculated in several
different ways. Let us use the following technique:
_p2j
р pref Ю2 ј _10Lpj
=10 ј 1:14439_109
_p2
j ј р20_10_6Ю2р1:14439_109Ю ј 0:45776 Pa2
The acoustic power for the motor may be calculated from Eq. (3-30):
W ј р9:817Юр0:45776Ю
р411:6Юр10Ю ј 1:092_10_3Wј 1:092mW
The sound power level for the motor is as follows:
LW ј 10 log10р1:092_10_3=10_12Ю ј 90:4dB
3.6.3 Sound Power Survey Measurement
There are some situations in which the noise source cannot be moved into a
reverberant room or into an anechoic room. If the noise source is not
located outdoors away from reflective surfaces, both the direct and diffuse
or reverberant acoustic fields will influence the relationship between sound
power and sound pressure. In this case, the microphone array location on a
rectangular parallelepiped shown in Fig. 3-11 may be used to estimate the
sound power from sound pressure measurements. The measured sound
pressure levels must be ‘‘corrected’’ for the presence of any background
noise, as discussed in Sec. 3.7.
The sound power level may be determined by the comparison method
using a calibrated sound power source, as discussed in Sec. 3.6.1.1. Equation
(3-8) may be used to calculate the sound power level LW for the source from
measurements of the sound pressure level Lp with the sound source in
operation, the sound pressure level Lp;cal with the calibrated sound power
source in operation alone, and the sound power level LW;cal given by the
manufacturer of the calibrated source.
62 Chapter 3
Copyright © 2003 Marcel Dekker, Inc.
The sound power may also be calculated from measurements of the
reverberation time for the space in which the noise source is located. If we
include the acoustic energy directly transmitted through the measurement
area Sm, Eq. (7-14) may be written in the following form:
W
4
R ю
1
Sm
_ _
ј
W
Sm
4Sm
R ю 1
_ _
ј
p2
av
_oc
(3-31)
The quantity pav is the energy-averaged sound pressure from the measurements:
p2
av ј
_p2j
Ns
(3-32)
The room constant R may be taken from Eq. (3-16), with the second term in
parenthesis neglected, since the magnitude of this term is usually smaller
than the uncertainty in the sound power determination:
W ј
Smp2
av
_oc
1 ю
cTrSm
13:816V
_ _
_1
(3-33)
Acoustic Measurements 63
FIGURE 3-11 Microphone locations for rectangular measuring surfaces, using 9
microphone locations. Point 1 is at the center of the top surface, points 2 through
5 are at the corners of the top surface, and points 6 through 9 are at the centers of the
vertical surfaces.
Copyright © 2003 Marcel Dekker, Inc.
We may convert Eq. (3-33) to ‘‘level’’ form by introducing the reference
quantities, then taking log base 10 of both sides and multiplying through by
10:
LW ј Lp;av ю10 log10рSm=SrefЮ_Kr _10 log10
_ocWref
p2
refSref
_ _
(3-34)
The reference area is Sref ј 1m2, and the quantity Kr is defined by the
following expression:
Kr ј 10log10 1 ю
cTrSm
13:816V
_ _
(3-35)
For air at 258C (298.2K or 778F) and 101.325 kPa (14.696 psia), the characteristic
impedance _oc ј 409:8 rayl. Using this value, we may determine
the numerical value of the last term in Eq. (3-34):
10log10
_ocWref
p2
refSref
_ _
ј 10log10 р409:8Юр10_12Ю
р20_10_6Ю2р1Ю
" #
ј 0:1 dB (3-36)
The sound power level expression may be written as follows, using the value
of 0.1 dB for the last term in Eq. (3-34):
LW ј Lp;av ю10 log10рSm=SrefЮ_Kr _0:1 (3-37)
There may be some situations in which the noise source cannot be
stopped or ‘‘turned off’’ in order that reverberation time measurements
can be made. In these cases, the sound power may be determined, with
some loss in accuracy, by first estimating the roomconstant. Using information
about the room surfaces and the techniques discussed in Chapter 7, the
room constant may be estimated from Eq. (7-13):
R ј
__S0
1___
(3-38)
The quantity __ is the average surface absorption coefficient for the room
surfaces, and S0 is the total surface area of the room in which the noise
source is located. The factor Kr in Eq. (3-34) is given by the following
expression, using the estimate for the average surface absorption coefficient:
Kr ј 10log10 1 ю
4р1___ЮSm
__S0
_ _
(3-39)
The rectangular parallelepiped measuring surface and the key measuring
points for the survey method of determining the sound power are illustrated
in Fig. 3-11 (ISO, 1986e). The reference surface, with dimensions
‘1; ‘2, and height ‘3, is the smallest parallelepiped that can enclose the noise
64 Chapter 3
Copyright © 2003 Marcel Dekker, Inc.
source. The measuring surface, on which the microphone measurement
points are located, has dimensions of р‘1 ю2dЮ; р‘2 ю2dЮ; and height
р‘3 юdЮ. The dimension d is somewhat flexible; however, a distance of
d ј 1m (39.4 in) is often used for cases in which the largest dimension
‘max of the reference surface is 250mm (9.8 in) or larger. For cases in
which the largest dimension of the reference surface is less than 250 mm,
the dimension d may be taken as any distance from р4‘maxЮ to 1m, but not
smaller than 250mm. For example, if the largest dimension of the reference
surface is 150mm (5.91in), the dimension d could be chosen as any value
from0.60m(23.6 in) to 1.00m(39.4in). However, if the largest dimension of
the reference surface is 50mm(1.97in), the dimension d would be chosen as
250mm (9.8 in) and not р4Юр50Ю ј 200mm, for example.
There are nine key microphone locations, including locations at the
height h ј 1
2 р‘3 юdЮ in the middle of the four vertical faces, at the center of
the top surface, and at each of the four corners of the top surface. Eight
additional microphone locations, including locations at the center of each of
the four edges of the top surface, and locations at the center of each of the
four vertical edges, may be used for additional accuracy.
Example 3-4. A small air compressor has envelope dimensions of 600mm
(23.6 in) wide by 800mm (31.5 in) long by 600mm (23.6 in) high. The compressor
is located in a room having dimensions of 15m (49.21 ft) by 18m
(59.06 ft) by 3.75m(12.30ft) high. The estimated average surface absorption
coefficient for the room is __ ј 0:15. The measurement surface is selected
with a spacing d of 1.00m (3.28ft or 39.4 in) from the compressor envelope
surfaces, such that the dimensions of the measurement surface are 2.60m
(8.53 ft) by 2.80m (9.19ft) by 1.60m (5.25ft) high. The measured sound
pressure level values are given in Table 3-3 for the 500Hz octave band.
Determine the sound power level of the compressor for the 500 Hz octave
band.
The energy-averaged sound pressure level may be determined from the
nine data points and Eq. (3-32):
рpav=pref Ю2 ј р108:20 ю 108:12 ю_ _ _ю107:81Ю=р9Ю ј 1:0972 _ 108
pav ј р20 _ 10_6Юр1:0972 _ 108Ю1=2 ј 0:2095 Pa
Lp;av ј 10 log10р1:0972 _ 108Ю ј 80:4dB
The measurement surface area is as follows:
Sm ј р2Юр2:60 ю 2:80Юр1:60Ю ю р2:60Юр2:80Ю
Sm ј 17:28 ю 7:28 ј 24:56m2 р264:4 ft2Ю
Acoustic Measurements 65
Copyright © 2003 Marcel Dekker, Inc.
The surface area of the roommay be determined. We may either neglect the
effect of the area covered by the compressor (less than 1% of the room
surface area) or we may include the surface acoustic absorption of the
compressor and exclude the floor area covered by the compressor. Let us
use the first approach, since the floor area covered by the compressor is
small:
So ј р2Юр15ю18Юр3:75Ююр2Юр15Юр18Ю
So ј 247:5ю540:0 ј 787:5m2 р8477 ft2Ю
The factor Kr may be calculated from Eq. (3-39) for this problem:
Kr ј 10 log10 1 ю р4Юр1_0:15Юр24:56Ю
р0:15Юр787:5Ю
_ _
ј 10 log10р1ю0:7069Ю
ј 2:32dB
The sound power level for the 500Hz octave band may be determined
from Eq. (3.37):
LW ј 80:4ю10log10р24:56Ю_2:32_0:1
LW ј 80:4ю13:90_2:33 ј 92:0 dB
3.6.4 Measurement of the Directivity Factor
The directivity factor Q or the directivity index DI, defined by Eqs (2-27)
and (2-28), may be determined frommeasurements of the sound power in an
anechoic or semi-anechoic room. The directivity may also be measured outdoors
far away from reflecting surfaces. If the measurement is made out-
66 Chapter 3
TABLE 3-3 Data for Example 3-4.
The measurement locations are
illustrated in Fig. 3-11
Point Location Lp, dB
1 Top, center 82.0
2 Top corner, front 81.2
3 Top corner, front 82.6
4 Top corner, back 79.6
5 Top corner, back 80.1
6 Vertical side, center 76.7
7 Vertical side, center 79.8
8 Vertical side, center 80.6
9 Vertical side, center 78.1
Copyright © 2003 Marcel Dekker, Inc.
doors, the microphone should be located at a distance such that the sound
pressure level decreases by 6 dB for each doubling of the distance from
source. Generally, the number of measurement points required for effective
determination of the directivity is larger than that needed for sound power
determination. The measurement ‘‘mesh’’ should be made finer in the
regions where the sound pressure varies rapidly with position.
The directivity factor Q is defined mathematically by Eq. (2-27). The
directivity factor is the ratio of the sound intensity in a specified direction to
the sound intensity for a spherical source having the same overall sound
power:
Q ј
I
рW=4_r2Ю ј
4_r2I
W р3-40)
The quantity r is the radial distance from the center of the source to the
point at which the intensity I is determined. The directivity index is the
directional characteristics expressed in ‘‘level’’ form, and is defined by Eq.
(2-28):
DI ј 10 log10рQЮ (3-41)
The acoustic power for the source must be measured first before the
directivity factor can be determined. Using measurements of the acoustic
pressure in an anechoic or semi-anechoic room on a spherical or hemispherical
surface of radius a, the directivity factor may be calculated from Eq. (3-
40), using Eqs (3-22) and (3-23) for the sound power:
Q ј
4_a2p2j
_oc_SjIj ј
4_a2p2j
_Sjp2j
(3-42)
If the total surface area is divided into Ns equal areas, Eq. (3-42) may be
simplified by making the substitutions from Eq. (3-24) or (3-25):
Q ј
2Nsp2j
_p2j
ј
2p2j
p2
av
(sound source on a reflective surface) (3-43)
Q ј
Nsp2j
_p2j
ј
p2j
p2
av
(sound source suspended freely) (3-44)
The quantity pav is the energy-averaged sound pressure obtained from the
sound power measurements:
p2
av ј
_p2j
Ns
(3-45)
Acoustic Measurements 67
Copyright © 2003 Marcel Dekker, Inc.
The directivity index may be found from Eq. (3-43) or (3-44) by introducing
the reference sound pressure рpref ј 20 mPa), taking log base 10 of
both sides of the equations, and multiplying through by 10:
DI ј Lpj _Lp;av ю3 dB (sound source on a reflective surface)
(3-46)
DI ј Lpj _Lp;av (sound source suspended freely) (3-47)
Example 3-5. Determine the directivity factor and directivity index for the
sound source given in Example 3-3 for an angle of _ ј 08 with the vertical
and for _ ј 41:48.
The square of the energy-averaged sound pressure is:
p2
av ј р_p2
j Ю=Ns ј р0:45776Ю=р10Ю ј 0:045776 Pa2
The measured acoustic pressure at _ ј 08 may be found from the data in
Table 3-2.
p1 ј р20 _ 10_6Юр1086=20Ю ј 0:3991 Pa
The measurements were taken in a semi-anechoic room, so Eq. (3-43)
may be used to evaluate the directivity factor:
Q ј р2Юр0:3991Ю2
р0:045776Ю ј 6:96 for _ ј 08
The directivity index is found from Eq. (3-41):
DI ј 10 log10р6:96Ю ј 8:43 dB for _ ј 08
We may use an alternative method to determine the directivity index. The
average sound pressure level is as follows:
Lp:av ј 10 log10Ѕр0:045776Ю=р20 _ 10_6Ю2_ ј 80:6dB
The directivity index may be calculated from Eq. (3-46) for measurements
taken in a semi-anechoic room:
DI ј 86:0 _ 80:6 ю 3 ј 8:4dB
From the data given in Table 3-2, we observe that the sound pressure
level at 41.48 does not vary more than about 1 dB with the angle ’. Let us
treat the source as an approximately symmetrical source. The average
acoustic pressure level is found by averaging the data from points 2, 3,
and 4:
68 Chapter 3
Copyright © 2003 Marcel Dekker, Inc.
Lp2;av ј 10 log10Ѕр1=3Юр108:15 ю 108:24 ю 108:13Ю_ ј 81:8dB
p2;av ј р20 _ 10_6Юр1081:8=20Ю ј 0:2461 Pa
We may use Eq. (3-43) to evaluate the directivity factor for _ ј 41:48:
Q_ ј р2Юр0:2461Ю2
р0:045776Ю ј 2:65 for _ ј 41:48
The directivity index for _ ј 41:48 is as follows:
DI_ ј 10 log10р2:65Ю ј 4:23 dB for _ ј 41:48
The directivity index could also be determined from Eq. (3-46):
Lp2;av ј 20 log10р0:2461=20 _ 10_6Ю ј 81:8dB
DI_ ј 81:8 _ 80:6 ю 3 ј 4:2dB
Популярные книги
- Старинные занимательные задачи
- Медоносные растения
- Algebratic geometry
- Workbook in Higher Algebra
- Математика Древнего Китая
- Finite element analysis
- Mathematics and art
- Fields and galois theory
- Пчеловодство
- Black Holes
Популярные статьи
- Higher-Order Finite Element Methods
- Электровакуумные приборы
- Riemann zeta functionS
- Универсальная открытая архитектурно-строительная система зданий серии Б1.020.1-71
- Complex Analysis 2002-2003
- Пример расчета прочности елементов, стыков и узлов несущего каркаса здания
- Составы, вещества и материалы для огнезащитыметаллических консрукций и изделий
- CMOS Technology
- Рекомендации по расчету и конструированию сборных железобетонных колонн каркасов зданий серии Б1.020.1-7 с плоскими стыками ВИНСТ
- Советы старого пчеловода